Daily maintenance: Clean the used RDA’s. Rinse and leave in soapy solution for several hours before drying. Hands-on: 10 minutes.
With 10 runs left (once a week): clean the metal plates with water and a brush, hands-on: 10 min.
After 20 runs on the most used station: perform DAB user cleaning on all stations. This is a run which takes 3 hours. Hands-on: 10 min.
Twice a month: perform bulk line cleaning. This takes 20 minutes total, hands-on: 5 min.